L-Amino Acid Quantitation Kit
- 貨號 Cat. No.
- K639-100
- 規格 Size
- 100 assays
Description
I. Introduction:
L-Amino acids are the most essential elements in biology. Accurately quantitating L-amino acids in body fluids or purified samples may provide valuable information for diagnostic or basic research studies. BioVision’s L-Amino Acid Assay Kit provides a convenient means for directly detecting L-amino acid in biological samples. There is no requirement for sample pretreatment or purification when using this kit. The L-amino acid(s) level can be quantified using fluorometric (at Ex/Em = 535/587 nm) or colorimetric (at l = 570 nm) methods in 96-well plates.
Kit Contents:
Components | K639-100 | Cap Code | Parts Number |
L-Amino Acid Assay Buffer L-Amino Assay Probe DMSO (Dried) L-Amino Acid Enzyme Mix L-Amino Acid Standard (4nmol/μl)* | 25 ml 1 Vial 500 μl 1 Vial 300μl | WM Red Brown Green Yellow | K639-100-1 K639-100-2 K639-100-3 K639-100-4 K639-100-5 |
![[image]](http://www.biovision.com/Merchant2/k639-100-l-amino-acid-quantitation-kit/k639-100-l-amino-acid-quantitation-kit-1.jpg)
*Mixture of all amino acids at equal molar ratio. Glycine can’t be detected by this kit.
Reagent Preparation and Storage Conditions:
Probe: Dissolve the probe with 220 μl of dried DMSO (provided) before use. Mix, store at –20C, protect from light and moisture. Use within two months.
Enzyme Mix: Dissolve in 220 μl L-Amino Acid Assay Buffer. Pipette up and down to complete dissolve the content. Store at –20C. Use within two months.
Assay Protocol:
1. Standard Curve Preparations: For colorimetric assay, add 0, 2, 4, 6, 8, 10 μl L-Amino Acid Standard into each well individually of a 96-well plate to generate 0, 8, 16, 24, 32, 40 nmol/well of L-Amino Acid Standard. Adjust volume to 50 μl/well with L-Amino Acid Assay Buffer.
For fluorometric assay, dilute the L-Amino Acid to 0.4 nmol/μl by adding 10 μl of the L-Amino Acid to 90 μl of L-Amino Acid Assay Buffer, mix well. Add 0, 2, 4, 6, 8, 10 μl into each well individually to generate 0, 0.8, 1.6, 2.4, 3.2, 4.0 nmol/well of the L-Amino Acid Standard. Adjust volume to 50 μl/well with L-Amino Acid Assay Buffer.
Sample Preparations: Prepare test samples in 50 μl/well with L-Amino Acid Assay Buffer in the 96-well plate. We suggest using several doses of your sample to ensure the readings are within the standard curve range.
3. Reaction Mix Preparation: Mix enough reagents for the number of assays performed: For each well, prepare a total 50 μl Reaction Mix containing:
46 μl L-Amino Acid Assay Buffer
2 μl L-Amino Acid Probe
2 μl L-Amino Acid Enzyme Mix,
Mix well.
Add 50 μl of the Reaction Mix to each well containing the L-Amino Acid standard or test samples.
Incubate the reaction for 30 minutes at 37oC, protect from light.
Measure O.D. 570nm for colorimetric assay or fluorescence at Ex/Em = 535/590 nm in a micro-plate reader.
Correct background by subtracting the value of the 0 L-amino acid control from all samples (The background reading can be significant and must be subtracted from sample readings). Then apply the sample readings to the L- amino acid standard curve to obtain the total amino acid amount.
L-Amino Acid Concentration = A/Sv (nmol/μl or mM)
Where A: L-Aamino acid amount (nmol) from the standard curve based on Absorbance OD570 or fluorescence of your samples.
Sv: Sample volume (μl) you added into the sample wells.
L-Amino Acid Standard Curve. Different doses of L-Amino Acids were measured according to the kit procedure.