Human Adiponectin ELISA Development Kit

貨號 Cat. No.
K4901-1000
規格 Size
10 x 96 wells assays

Description

 









(Catalog #K4901-100; Store kit at +4ºC)


I. Description:


Adipose tissue secretes a number of biologically active soluble factors (collectively named adipocytokines) that regulate glucose and fatty acid metabolism. Measurement of serum adiponectin levels gives us important information on the role of adiponectin in regulation of glucose and/or lipid metabolism. This human Adiponectin ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for quantitative determination of adiponectin in human serum, plasma or various tissue or cell culture supernatants. In the assay, polyclonal antibody specific for mouse adiponectin has been pre-coated onto 96 well microplate. Standards and samples are pipetted into the wells and adiponectin present is bound by immobilized antibody. The bound adiponectin is then captured by anti-rat adiponectin monoclonal antibody. With HRP conjugated mouse IgG and a HRP substrate, the colors developed in proportion to the bound adiponectin, can be easily measured by Elisa plate reader. 




  1. Kit Components:






  1. Antibody coated 96-well plate (12 x 8-well strips)




  2. 5X Wash concentrate, 100 ml




  3. 5X Diluent, 50 ml




  4. Secondary antibody, 12 ml




  5. 100X Detector, 150 µl




  6. Standard, recombinant human adiponectin (64 ng), 1 vial, lyophilized




  7. QC sample= positive control having 7-11 µg/ml human adiponectin, 1 vial, lyophilized




  8. Substrate I, 6 ml




  9. Substrate II, 6 ml




  10. Stop solution, 12 ml




 




  1. Storage Conditions:




Reagents must be stored at 2-8ºC when not in use. The reagents must be brought up to room temperature before use. Do not expose the reagents to temperature above 25ºC. Diluted wash solution may be stored at room temperature for up to one month. 


IV. Assay Procedure


A. Preparation of Reagents


1. Allow all samples and kit components to equilibrate to room temperature (20-25°C).


2. Plan the plate configuration and create a plate map. Calculate the amount of working reagents to use (See table below).


It is recommended that standards and samples be run in duplicate.


3. Prepare 1X Wash Solution: Dilute 5X Wash Concentrate to 1X with deionized water. The diluted 1X Wash Solution is stable for one month at room temprature


4. Prepare 1X Diluent. Dilute 5X Diluent to 1X with deionized water.


5. Prepare 1X Detector. Dilute 100X Detector to 1X with 1X Diluent. Use the 1X Detector within one hour of preparation.





    1. Prepare Substrate Solution freshly by adding one part Substrate I to one part Substrate II. Freshly prepare just before use.





 


 


 


 


 









The amount of working reagents needed for 1 well









































Working

reagents


Total

volume needed


Stock

solution added


Dilution

solution added


Note


1X Wash Solution


2.8 ml


0.56 ml of 5X Wash Concentrate


2.24 ml of ddH2O


Stable for 1 month at RT


1X Diluent


2.5 ml


0.5 ml of

5X Diluent


2.0 ml of ddH2O


in the case of 10 µl sample;

Including standard dilution


1X Detector


110 µl


1.1 µl of 100X Detector


108.9 µl of 1X Diluent


Use within 1 hr.


Substrate

Solution


110 µl


55 µl of

Substrate I


55 µl of

Substrate II


Freshly prepared just

before use


 


7. Prepare working aliquots of the Standard as follows:


Briefly centrifuge the lyophilzed Standard vial. When opening, remove cap gently as the lyophilizate may have become dislodged during shipping. Add 1 ml of deionized water the Standard vial to make a stock concentration of 64 ng/ml. Mix well. A recommended dilution scheme is as follows:


1) Label 8 microcentrifuge tubes #0-7. Add 200 µl Diluent to the microcentrifuge tubes #0-7.


2) Add 200 µl of the stock Standard solution to tube #7 and vortex. This is Standard tube #7 with a concentration of 32 ng/ml




  1. Standards #6 to #1 are then prepared by performing a 1:2 dilution of the preceding standard. Do not add any standard to the tube #0




 


300 µl 300 µl 300 µl 300 µl 300 µl 300 µl 300 µl[image][image][image][image][image][image][image][image][image][image][image][image][image][image][image]


 


 


 


Stock


Standard


 


Tube # stock 7 6 5 4 3 2 1 0


Conc.(ng/ml) 64 32 16 8 4 2 1 0.5 0


Initial vol.(µl) - 300 300 300 300 300 300 300 300


 


 


8. Reconstitute QC sample in 1 ml of deionized water. 


 


 


 


 


 


 


 


B. Sample dilution


Step 1. Dilute samples 1:100 with 1X Diluent (for example, 10 ul sample plus 990 ul 1X Diluent, final 1:100)


Step 2. Dilute the samples (from step 1) 1:20 with 1X Diluent (for example, 50 µl step 1 sample plus 950 µl 1X Diluent, final 1:2000)


* If samples fall the outside range of assay, a lower or higher dilution may be required.


Step3. Use 100 ul of the final sample for ELISA.


 


 



 




50 µl




Use 100 µl for assay



[image][image]

Add 10 ul sample to


990 µl 1X Diluent




Dilution factor


= 100




Dilution factor


= 2000



[image]

950 µl


1X Diluent



[image]

 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


C. Experiment procedure


1. Remove the appropriate number of microwell strips from the sealed foil pouch.


2. Pipette 100 µl of standards 0 to 7, the reconstituted QC sample and pre-treated plasma sample into the antibody-coated plate according to the plate configuration. Use a new pipette tip for each standard or sample.


3. Incubate at 37°C for 1 hour.


4. Remove the solution and wash 3 times with 250 µl of 1X Wash Solution to each well.


5. Add 100 µl Secondary Antibody to each well.


6. Incubate at 37°C for 1 hour.


7. Remove the solution and wash 3 times with 250 µl of 1X Wash Solution to each well.


8. Add 100 µl 1X Detector to each well.


9. Incubate at 37°C for 1 hour.


10. Remove the solution and wash 5 times with 250 µl of 1X Wash Solution to each well.


11. Using the multi-channel pipette, add 100 µl of the Substrate Solution to each well.


12. Incubate at room temperature for 20 min. Protect from light.


13. Using the multi-channel pipette, add 100 µl Stop Solution to each well.


14. Read at 450 nm.


15. Subtract the absorbance of the blank from the readings for each standard and sample.




  1. Construct a standard curve by plotting the known concentrations (Y) of standard versus the absorbances (X) of standard. A measurable range is typically shown between 0.5 ng/ml and 32 ng/ml.




 


 


 


 


 


17. Calculate the adiponectin concentrations of samples by interpolation of the regression curve formula as shown above in a form of a quadratic equation.


18. The adiponectin concentrations calculated must be multiplied by dilution factor to obtain the concentrations of the undiluted samples (Dilution factor of lyophilized QC sample is 2000) 


[image]


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 




  1. Performance Characteristics:




    1. Sensitivity: The limit of detection: 100 pg/ml.




    2. Specificity: No cross-reaction with mouse and rat adiponectin.




    3. Recovery: The average recovery of adiponectin is 90-105%.