Glucose and Sucrose Assay Kit
- 貨號 Cat. No.
- K616-100
- 規格 Size
- 100 assays
Description
(Catalog #K616-100; 100 assays; Store at -20oC) I. Introduction: Glucose (C6H12O6; FW: 180.16) and sucrose (C12H22O11; FW:342.3) are the important fuel sources to generate universal energy molecule ATP. Measurement of glucose or sucrose level can be very important in both research and development process. Sucrose is a disaccharide which can be converted into one glucose and one fructose when adding Invertase. BioVision’s Glucose and Sucrose Assay Kit provides a convenient means for measuring glucose and sucrose levels from various biological samples (e.g. serum, plasma, body fluids, food, growth medium, etc.). To measure glucose level, glucose oxidase specifically oxidizes free-glucose generating a compound that reacts with the glucose probe to produce resorufin, which can be detected colorimetrically (O.D. 570 nm) or fluorometrically (Ex/Em 535/587). To measure sucrose, invertase can be added to the reaction to convert sucrose to free glucose and fructose, so total glucose level can be measured. Then the sucrose level = Total Glucose – Free Glucose.
Store kit at –20oC, protect from light. Allow reagents warm to room temperature and briefly centrifuge vials before opening.
Glucose Probe: Dissolve in 220 μl DMSO (provided) before use. Store at –20oC, protect from light and moisture. Use within two months. Invertase: Dissolve in 220 μl Glucose Assay Buffer. Aliquot and store at –20oC. Use within two months. Glucose Enzyme Mix: Dissolve in 220 μl Glucose Assay Buffer. Aliquot and store at –20oC. Use within two months.
1. Standard Curve Preparations: For colorimetric assay, dilute the Sucrose Standard to 1 nmol/μl by adding 10 μl of the Sucrose Standard to 990 μl of Glucose Assay Buffer, mix well. Add 0, 2, 4, 6, 8, 10 μl into each well individually. Adjust volume to 50 μl/well with Glucose Assay Buffer to generate 0, 2, 4, 6, 8, 10 nmol/well of Sucrose Standard.
For fluorometric assay, dilute the Sucrose Standard solution to 0.1 nmol/μl by adding 10 μl of the Sucrose Standard to 990 μl of Glucose Assay Buffer, mix well. Then take 20 μl into 180 μl of Glucose Assay Buffer. Mix well. Add 0, 2, 4, 6, 8, 10 μl into each well individually. Adjust volume to 50 μl/well with Glucose Assay Buffer to generate 0, 0.2, 0.4, 0.6, 0.8, 1.0 nmol/well of the Sucrose Standard. Fluorometric assay is 10-100 fold more sensitive than colorimetric assay.
Note: 2 μl of invertase must be added to each well of the Sucrose Standard to convert sucrose standard to glucose for either glucose or sucrose assay.
46 μl Glucose Assay Buffer 2 μl Glucose Probe 2 μl Glucose Enzyme Mix
|