Annexin V-PE Apoptosis Kit Plus
- 貨號 Cat. No.
- K203-25
- 規格 Size
- 25 assays
Description
(Catalog #: K203-25, -100, -400; Store kit at 4oC) I. Introduction: The Annexin V-PE Apoptosis Detection Kit Plus is based on the observation that soon after initiating apoptosis, most cell types translocate the membrane phospholipid phosphatidylserine (PS) from the inner face of the plasma membrane to the cell surface. Once on the cell surface, PS can easily be detected by staining with a fluorescent conjugate of Annexin V, a protein that has a strong natural affinity for PS. The one-step staining procedure takes only 10 minutes. In addition, the assay can be directly performed on live cells, without the need of fixation. The Annexin V-PE Apoptosis Detection Kit Plus includes annexin V-PE, SYTOX green dye, and binding buffer. The SYTOX green dye is impermeable to live cells and apoptotic cells, but stains necrotic cells with intense green fluorescence by binding to cellular nucleic acids. After staining a cell population with annexin V-PE and SYTOX Green dye in the provided binding buffer, apoptotic cells show red fluorescence, dead cells show green fluorescence and live cells show little or no fluorescence. These populations can easily be distinguished by Fluorescence microscopy using FITC and PE filters or by flow cytometry using the FL1 channel (Ex. 488 nm/Em. 530 nm) for SYTOX Green dye and FL2 channel for Annexin V-PE (Ex. 488 nm/Em. 578 nm).
Components K203-25 K203-100 K203-400 25 assays 100 assays 400 assays
Annexin V-PE 125 μl 500 μl 2 ml SYTOX Green Dye 25 μl 100 μl 400 μl Binding Buffer 12.5 ml 50 ml 2 x 100 ml
1. Induce apoptosis by desired method. Concurrently incubate a control culture without induction. 2. Collect 1-5 x 105 cells by centrifugation. 3. Resuspend cells in 500 µl of 1X Binding Buffer.
Note: Thaw the SYTOX Green dye in room temperature before use. 5. Incubate at room temperature for 5-10 min in the dark.
The cell population should separate into three groups: live cells with only a low level of fluorescence, apoptotic cells with red fluorescence and necrotic cells with green fluorescence. The flow cytometric results can also be confirmed by viewing the cells under a fluorescence microscope using FITC filter for SYTOX and PE filter for Annexin V-PE. For adherent cells, gently trypsinize and wash cells once with serum-containing media before incubation with Annexin V-PE and SYTOX dye.
Store kit at 4oC. All reagents are stable for one year under proper storage conditions |